Oxygen Index Tester Repair and Maintenance Shandong

Oxygen index meter repair and maintenance

Qingdao Zhongbang Instrument Co., Ltd. Equipment Technical Support

  First, the instrument fault diagnosis and processing methods

malfunction

the reason

Approach

The pipeline is not ventilated

1, N2 or O2 cylinder total valve, pressure relief valve is not open or the pressure is not enough

2, N2 or O2 regulator valve open in the end

Open and adjust

Reconditioning

Flow control valve screw does not advance

Screw thread

Replace the screw

The instrument has a loud sound after ventilation

N2 or O2 regulator valve spring shift

Adjustment

Digital display can not reach fullness

Sensor life exceeded

Replace sensor

Second, the instrument maintenance 1, the instrument if not used for a long time, it is recommended to remove the oxygen sensor, into the refrigerator freezer to extend the life of the sensor (detailed removal steps see step three).

2. Please read the user's manual and oxygen index test related national standards before opening the instrument, and be familiar with the correct use of the oxygen index meter and the calculation of the oxygen index value.

3, the test sample, the remaining part of the well-preserved, so that the value of the instrument in the future comparison test to determine the stability of the instrument, if more than the original accuracy level, should consider replacing the oxygen concentration sensor.

Third, the replacement of oxygen sensor 1, cut off the power and air supply.

2, remove the screws on both sides of the panel, grab the "regulator" valve knob and pull it out.

3. Unplug the three-pin connector on the center of the back panel of the instrument, remove the oxygen sensor counterclockwise (the face is a pentagonal shape), replace it with a new sensor, and insert the plug and reinstall the panel.

4, can be used after correcting fullness.

Fourth, note 1, the use of the specified voltage, instrument power should have a ground wire.

2. When testing samples, some materials release toxic and harmful gases. It is recommended that the test should be conducted in a fume hood and be well-protected.

3. When using liquefied gas, natural gas, gas and other gas sources, the pipeline interface should not be leaky, and the aging of the ventilation pipe should be promptly replaced.

4, the operator can not leave the test site during the test.

5, computer-controlled equipment should prevent the invasion of the virus.

6, the laboratory equipped with fire fighting equipment.

7. Close all power and gas sources at the end of the test.

660nm LED Grow Light

Ultra Plantâ„¢ Grow Light offers One Chip Technology aimed to meet your indoor growing expectation such as improve plants' quality, increase yield, or better the margin, etc., all for helping you realize a higher return on your crops.


Ultra Plantâ„¢ Grow Light is combined our advanced All-In-One technology with patented optical design and customized light full spectrum supported from our experienced LED engineers, plant specialists and other partners working on horticulture.


From Ultra Plantâ„¢ APP, you are able to schedule the growing process including photoperiod, brightness and spectral in advance. The lighting system will help you grow smarter, easier and better.


Ultra Plantâ„¢ is the most versatile horticultural grow lighting fixture for indoor plants with flexible full spectrum, brightness control and uniform, wider light distribution, suitable for top lighting of all types of crops. No matter it applies to anywhere for any crop, Ultra Plantâ„¢ can do perfect work for you.

660Nm Led Grow Light,660Nm Grow Lights,660Nm Grow Lights,660Nm Fluorescent Grow Lights

Feton Corporation , https://www.fetonledlight.com

Posted on